DocumentCode :
3227494
Title :
Angular spectrum approach for fast simulation of pulsed non-linear ultrasound fields
Author :
Du, Yigang ; Jensen, Henrik ; Jensen, Jørgen Arendt
Author_Institution :
Dept. of Electr. Eng., Tech. Univ. of Denmark, Lyngby, Denmark
fYear :
2011
fDate :
18-21 Oct. 2011
Firstpage :
1583
Lastpage :
1586
Abstract :
The paper presents an Angular Spectrum Approach (ASA) for simulating pulsed non-linear ultrasound fields. The source of the ASA is generated by Field II, which can simulate array transducers of any arbitrary geometry and focusing. The non-linear ultrasound simulation program - Abersim, is used as the reference. A linear array transducer with 64 active elements is simulated by both Field II and Abersim. The excitation is a 2-cycle sine wave with a frequency of 5 MHz. The second harmonic field in the time domain is simulated using ASA. Pulse inversion is used in the Abersim simulation to remove the fundamental and keep the second harmonic field, since Abersim simulates non-linear fields with all harmonic components. ASA and Abersim are compared for the pulsed fundamental and second harmonic fields in the time domain at depths of 30 mm, 40 mm (focal depth) and 60 mm. Full widths at -6 dB (FWHM) are {0.97, 0.95} mm at the focal depth for the fundamental fields for ASA and Abersim, and {0.56, 0.55} mm for the second harmonic fields. Full widths at -12 dB are {1.27, 1.26} mm for the fundamental fields for ASA and Abersim, and {0.77, 0.73} mm for the second harmonic fields. The calculation time, for the second harmonic fields, using ASA is 12 minutes and for all harmonic fields using Abersim is 14 hours. Compared to Abersim, the error of ASA for calculating the second harmonic fields is 1.5% at -6 dB and 6.4% at -12 dB, and the calculation time is reduced by a factor of 70.
Keywords :
acoustic field; harmonics; time-domain analysis; ultrasonic focusing; ultrasonics; ASA; Abersim simulation; angular spectrum approach; focusing; frequency 5 MHz; linear array transducer; pulse inversion; pulsed nonlinear ultrasound fields; second harmonic field; time domain; Acoustics; Arrays; Harmonic analysis; Imaging; Mathematical model; Transducers; Ultrasonic imaging;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2011 IEEE International
Conference_Location :
Orlando, FL
ISSN :
1948-5719
Print_ISBN :
978-1-4577-1253-1
Type :
conf
DOI :
10.1109/ULTSYM.2011.0393
Filename :
6293278
Link To Document :