Title :
TOHR: prototype design and characterization of an original small animal tomograph
Author :
Ploux, L. ; Mastrippolito, R. ; Pinot, L. ; Laniéce, P. ; Charon, Y. ; Ochoa, A. Valda ; Valentin, L.
Author_Institution :
Inst. de Phys. Nucl., Univ. de Paris 11, Orsay, France
Abstract :
The authors have proposed an original method for high resolution in vivo tomography adapted to small animal models studies: TOHR (French acronym for High Resolution Tomograph). They have designed and developed a large solid angle, high resolution and high efficiency focusing collimator which provides a focal point used to scan the animal and finally perform the image. High resolution was improved by using nuclides having a two-photon decay (X- or gamma rays). The authors have built a prototype of a tomograph with a 2π collimator (10 over the 20 sides of an icosahedron). The different parts of the detector (collimator, detector, acquisition) are described. An optimization of the collimator structure and an iso-focusing method of the different modules of the collimator are discussed. Performances of the detector (energy, timing and spatial resolution) are presented using point sources and hot spot phantoms and are compared to numerical simulation results
Keywords :
biological techniques; emission tomography; 2π collimator; TOHR; animal research; collimator structure; energy; high resolution high efficiency focusing collimator; hot spot phantoms; icosahedron; isofocusing method; large solid angle collimator; numerical simulation results; point sources; prototype design; small animal tomograph characterization; spatial resolution; timing; two-photon decay; Animals; Collimators; Detectors; Focusing; Gamma rays; Image resolution; In vivo; Prototypes; Solids; Tomography;
Conference_Titel :
Nuclear Science Symposium, 1997. IEEE
Conference_Location :
Albuquerque, NM
Print_ISBN :
0-7803-4258-5
DOI :
10.1109/NSSMIC.1997.670493