DocumentCode :
322824
Title :
Comparison of multi-pole shaping and delay line clipping pre-amplifiers for position sensitive NaI(Tl) detectors
Author :
Freifelder, R. ; Karp, J.S. ; Wear, J.A. ; Lockyer, N.S. ; Newcomer, F.M. ; Surti, S. ; Van Berg, R.
Author_Institution :
Dept. of Radiol., Pennsylvania Univ., Philadelphia, PA, USA
Volume :
2
fYear :
1997
fDate :
9-15 Nov 1997
Firstpage :
1170
Abstract :
NaI(Tl) position sensitive detectors have been used in medical imaging for many years. For PET applications without collimators, the high counting rates place severe demands on such large area detectors. The NaI(Tl) detectors in the PENN-PET scanners are read-out via photomultiplier tubes and preamplifiers. Those preamplifiers use a delay-line clipping technique to shorten the characteristic 240 ns fall time of the NaI(Tl) signal. As an alternative, the authors have investigated a pole-zero network to shorten the signal followed by a multi-pole shaper to produce a symmetric signal suitable for high counting rates. This has been compared to the authors´ current design by measuring the energy and spatial resolution of a single detector as a function of different preamplifier designs. Data were taken over a range of ADC integration times and countrates. The new design shows improved energy resolution with short integration times. Effects on spatial resolution and deadtime are reported for position sensitive detectors at different countrates
Keywords :
biomedical electronics; biomedical equipment; image resolution; photomultipliers; position sensitive particle detectors; positron emission tomography; sodium compounds; solid scintillation detectors; thallium; 240 ns; NaI:Tl; PENN-PET scanners; PET applications without collimators; delay line clipping pre-amplifiers; medical diagnostic imaging; medical instrumentation; multi-pole shaping; nuclear medicine; photomultiplier tubes; position sensitive NaI(Tl) detectors; Biomedical imaging; Collimators; Current measurement; Delay effects; Delay lines; Photomultipliers; Position sensitive particle detectors; Positron emission tomography; Preamplifiers; Spatial resolution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium, 1997. IEEE
Conference_Location :
Albuquerque, NM
ISSN :
1082-3654
Print_ISBN :
0-7803-4258-5
Type :
conf
DOI :
10.1109/NSSMIC.1997.670518
Filename :
670518