Title :
Real-time rectilinear volumetric acquisition with a 7.5 MHz dual-layer array transducer - Data acquisition and signal processing
Author :
Chen, Yuling ; Nguyen, Man ; Yen, Jesse T.
Author_Institution :
Dept. of Biomed. Eng., Univ. of Southern California, Los Angeles, CA, USA
Abstract :
In previous work, we successfully developed a 7.5 MHz dual-layer array transducer for 3-D rectilinear imaging. This transducer significantly reduces the fabrication complexity and the channel count compared to fully sampled 2-D arrays, making 3-D imaging more realizable. Also, higher frequency (>; 5 MHz) 3-D imaging can provide more accurate volume measurements, detection of cystic or cancerous masses, and also assist in 3-D ultrasound guided biopsy and carotid imaging. In this paper, we present our work toward real-time 3-D imaging with this transducer, including real-time data acquisition with the Verasonics data acquisition system (VDAS), offline beamforming and display for volumetric images. The volume rate was set as 30 volume/sec with a field of view equals to 19.2 × 19.2 × 42.2 mm3. Movies were created to present a real-time display of the moving phantoms of multi-wire targets or cyst targets.
Keywords :
data acquisition; signal processing; 3D imaging; Verasonics data acquisition system; dual-layer array transducer; offline beamforming; real-time rectilinear volumetric acquisition; signal processing; volume measurements; Array signal processing; Arrays; Data acquisition; Phantoms; Real time systems; Transducers; 3-D; dual-layer; real-time data acquisition;
Conference_Titel :
Ultrasonics Symposium (IUS), 2011 IEEE International
Conference_Location :
Orlando, FL
Print_ISBN :
978-1-4577-1253-1
DOI :
10.1109/ULTSYM.2011.0439