Title :
Tissue harmonic imaging with CMUTs
Author :
Legros, M. ; Novell, A. ; Bouakaz, A. ; Férin, G. ; Dufait, R. ; Certon, D.
Author_Institution :
Vermon, Tours, France
Abstract :
In this work, we report on the characterization of a CMUT probe (Capacitive Micromachined Ultrasound Transducer) for Tissue Harmonic Imaging (THI). The intrinsic nonlinear behavior of the CMUT probe was first investigated. Matched electrical waveforms were transmitted to limit the impact of the transmit response distortion. With the implemented method, we demonstrated higher performances through in-vitro harmonic imaging.
Keywords :
capacitive sensors; distortion; micromachining; ultrasonic imaging; ultrasonic transducers; CMUT probe; THI; capacitive micromachined ultrasound transducer; distortion; electrical waveforms; in-vitro harmonic imaging; tissue harmonic imaging; Acoustics; Harmonic analysis; Imaging; Power harmonic filters; Probes; Transducers; Ultrasonic imaging; CMUTs; Tissue Harmonic Imaging;
Conference_Titel :
Ultrasonics Symposium (IUS), 2011 IEEE International
Conference_Location :
Orlando, FL
Print_ISBN :
978-1-4577-1253-1
DOI :
10.1109/ULTSYM.2011.0558