DocumentCode
3233035
Title
Accurate femoral canal shaping in total hip arthroplasty using a mini-robot
Author
Yoon, Yong Soo ; Lee, John Jaehwan ; Won, C.H. ; Hodgson, Antony J.
Author_Institution
Korea Adv. Inst. of Sci. & Technol., Seoul, South Korea
Volume
4
fYear
2001
fDate
2001
Firstpage
3214
Abstract
We propose a greatly simplified femoral canal shaping method for robot-assisted total hip replacement surgery that requires neither CT/MRI imaging nor insertion of fiducial markers before surgery. A reamer-shaped registration gauge is inserted into the femoral canal and used to position a base for a robot that is mounted on the top of the femur and used to prepare the femoral canal. In a study using 5 fresh cadavers, the average registration error for selected points near the surface of the femoral cavity was 0.59 mm; this is comparable to the errors reported for the geometry-based registration methods.
Keywords
medical robotics; surgery; femoral canal shaping; gauge based registration; hip implant; hip replacement; medical robotics; mini-robot; surgery; total hip arthroplasty; Bones; Computational geometry; Computed tomography; Hip; Implants; Irrigation; Magnetic resonance imaging; Pins; Robot kinematics; Surgery;
fLanguage
English
Publisher
ieee
Conference_Titel
Robotics and Automation, 2001. Proceedings 2001 ICRA. IEEE International Conference on
ISSN
1050-4729
Print_ISBN
0-7803-6576-3
Type
conf
DOI
10.1109/ROBOT.2001.933113
Filename
933113
Link To Document