A comparison of CMUTs and piezoelectric transducer elements for 2D medical imaging based on conventional simulation models
Author :
Wygant, I.
Author_Institution :
Adv. Technol. Group, Texas Instrum., Santa Clara, CA, USA
fYear :
2011
fDate :
18-21 Oct. 2011
Firstpage :
100
Lastpage :
103
Abstract :
This work compares CMUT and piezoelectric transducer technology for a typical 1D 7.5-MHz medical transducer array element. The comparison is based on the Redwood model of the piezoelectric transducer and a spring-mass-damper model for the CMUT. The simulated CMUT shows over twice the fractional bandwidth of the piezoelectric transducer but one-half the sensitivity and lower noise-equivalent-voltage. Experimental data from a CMUT array element generally confirms the CMUT simulations although shows lower source pressure and bandwidth.
Keywords :
biomedical transducers; biomedical ultrasonics; capacitive sensors; micromachining; piezoelectric transducers; ultrasonic transducer arrays; 1D medical transducer array element; 2D medical imaging; CMUT array element; CMUT simulations; Redwood model; capacitive micromachined ultrasound transducer; fractional bandwidth; noise-equivalent-voltage; piezoelectric transducer elements; piezoelectric transducer technology; simulated CMUT; simulation models; source pressure; spring-mass-damper model; Bandwidth; Sonar equipment; Transducers;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Ultrasonics Symposium (IUS), 2011 IEEE International
Conference_Location :
Orlando, FL
ISSN :
1948-5719
Print_ISBN :
978-1-4577-1253-1
Type :
conf
DOI :
10.1109/ULTSYM.2011.0025
Filename :
6293606
Link To Document :
بازگشت