    A comparison of CMUTs and piezoelectric transducer elements for 2D medical imaging based on conventional simulation models

  • Author

    Wygant, I.

  • Author_Institution
    Adv. Technol. Group, Texas Instrum., Santa Clara, CA, USA
  • fYear
    2011
  • fDate
    18-21 Oct. 2011
  • Firstpage
    100
  • Lastpage
    103
  • Abstract
    This work compares CMUT and piezoelectric transducer technology for a typical 1D 7.5-MHz medical transducer array element. The comparison is based on the Redwood model of the piezoelectric transducer and a spring-mass-damper model for the CMUT. The simulated CMUT shows over twice the fractional bandwidth of the piezoelectric transducer but one-half the sensitivity and lower noise-equivalent-voltage. Experimental data from a CMUT array element generally confirms the CMUT simulations although shows lower source pressure and bandwidth.
  • Keywords
    biomedical transducers; biomedical ultrasonics; capacitive sensors; micromachining; piezoelectric transducers; ultrasonic transducer arrays; 1D medical transducer array element; 2D medical imaging; CMUT array element; CMUT simulations; Redwood model; capacitive micromachined ultrasound transducer; fractional bandwidth; noise-equivalent-voltage; piezoelectric transducer elements; piezoelectric transducer technology; simulated CMUT; simulation models; source pressure; spring-mass-damper model; Bandwidth; Sonar equipment; Transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium (IUS), 2011 IEEE International
  • Conference_Location
    Orlando, FL
  • ISSN
    1948-5719
  • Print_ISBN
    978-1-4577-1253-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2011.0025
  • Filename
    6293606