DocumentCode
3235398
Title
Capon beamforming applied to second-harmonic ultrasound experimental data
Author
Näsholm, Sven Peter ; Austeng, Andreas ; Jensen, Are Charles ; Nilsen, Carl-Inge Colombo ; Holm, Sverre
Author_Institution
Dept. of Inf., Univ. of Oslo, Oslo, Norway
fYear
2011
fDate
18-21 Oct. 2011
Firstpage
2217
Lastpage
2220
Abstract
This paper presents the novelty of adaptive Capon beamforming on received experimental second-harmonic pulse-echo ultrasound imaging channel-data. IQ channel datasets were collected when imaging a tissue-mimicking phantom: two with the transmit frequency 1.7 MHz and the receive band-pass filters centered around 1.7 and 3.4 MHz respectively, as well as one with both the transmit and receive center frequencies being 3.4 MHz. Images are displayed and analyzed for all datasets both using Capon and conventional delay-and-sum beamforming. This work proves the feasibility of Capon beamforming applied to tissue-harmonic imaging.
Keywords
acoustic filters; acoustic signal processing; array signal processing; band-pass filters; biological tissues; biomedical ultrasonics; biomimetics; cellular biophysics; echo; phantoms; ultrasonic devices; ultrasonic imaging; ultrasonic transmission; IQ channel datasets; adaptive Capon beamforming; band-pass filters; center frequencies; conventional delay-and-sum beamforming; frequency 1.7 MHz; frequency 3.4 MHz; frequency transmission; second-harmonic pulse-echo ultrasound imaging channel-data; tissue-harmonic imaging; tissue-mimicking phantom; Acoustics; Array signal processing; Arrays; Biomedical imaging; Frequency conversion; Ultrasonic imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium (IUS), 2011 IEEE International
Conference_Location
Orlando, FL
ISSN
1948-5719
Print_ISBN
978-1-4577-1253-1
Type
conf
DOI
10.1109/ULTSYM.2011.0550
Filename
6293686
Link To Document