Abstract :
A simple and wideband calibration method for calibrating electric field sensors and probes with traceability to national standards using a GTEM cell and a transfer standard is investigated. We also introduce a method for the isotropic response measurement for electric field probes. A correction way for calibrating combined electric field strength in GTEM cell. A comparison with TEM cells is made to validate the proposed method.
Keywords :
TEM cells; calibration; electric field measurement; electric sensing devices; GTEM cell; electric field probes; electric field sensor calibration; field strength measurement standard; isotropic response measurement; transfer standard; wideband calibration method; Anechoic chambers; Calibration; Electric variables measurement; Electronic equipment testing; Horn antennas; Measurement standards; Metrology; Probes; TEM cells; Wideband;