• DocumentCode
    3244725
  • Title

    Development of a solid Xe ionization chamber

  • Author

    Nawa, H. ; Tamagawa, Y. ; Miyajima, M.

  • Author_Institution
    Dept. of Appl. Phys., Fukui Univ., Japan
  • fYear
    1999
  • fDate
    1999
  • Firstpage
    316
  • Lastpage
    318
  • Abstract
    Liquid Xenon (LXe) has good properties required for radiation detector application, such as small W-value and W5-value, large electron drift velocities, and can be operated in ionization and scintillation mode. From the results so far reported, solid xenon (SXe) has similar properties to that of LXe and the electron drift velocity in SXe is faster than that of LXe. We have recognized the potential of xenon for radiation detector media in the solid phase to be as high as that of LXe. In this paper, we discuss the properties of SXe from the view point of radiation detection and propose a solid ionization chamber for studying these characteristics
  • Keywords
    electron mobility; ionisation chambers; solid scintillation detectors; xenon; W-value; W5-value; Xe; electron drift velocity; gridded ionization chamber; radiation detector; radiation detector media; scintillation counters; solid Xe ionization chamber; Electron mobility; Energy resolution; Gases; Ionization chambers; Physics; Radiation detectors; Solids; Transistors; Velocity measurement; Xenon;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Dielectric Liquids, 1999. (ICDL '99) Proceedings of the 1999 IEEE 13th International Conference on
  • Conference_Location
    Nara
  • Print_ISBN
    0-7803-4759-5
  • Type

    conf

  • DOI
    10.1109/ICDL.1999.798934
  • Filename
    798934