Title :
Calibration of Optical Probe for CMM
Author :
Liu Shugui ; Tang Shaliang ; Chen Chao
Author_Institution :
State Key Lab. of Precision Meas. Technol. & Instrum., Tianjin Univ., Tianjin, China
Abstract :
A posture error compensation method of optical probe, which is installed with a PH10 rotating head for the scanning measurement of coordinate measuring machine (CMM), is presented in this paper. The error compensation mathematical model is derived from the least square fitting method and an optimization algorithm is applied to determine the optimal parameters. Through the experiments, the arithmetic´s accuracy and the calibration effectiveness of optical probe are proved.
Keywords :
calibration; coordinate measuring machines; error compensation; least squares approximations; optimisation; PH10 rotating head; calibration; coordinate measuring machine; least square fitting; optical probe; optimization; posture error compensation; scanning measurement; Arithmetic; Calibration; Coordinate measuring machines; Error compensation; Least squares methods; Magnetic heads; Mathematical model; Optimization methods; Probes; Rotation measurement;
Conference_Titel :
Photonics and Optoelectronics, 2009. SOPO 2009. Symposium on
Conference_Location :
Wuhan
Print_ISBN :
978-1-4244-4412-0
DOI :
10.1109/SOPO.2009.5230075