DocumentCode
3257409
Title
Effect study of bridge voltage on measurement of thermal conductivities of liquids by thermal probe method
Author
Zhang, Min ; Lu, Jiahua ; Chen, Jianhua ; Che, Zhenhua ; Le Yang ; Zhong, Zhiyou ; Zhao, Huizhong
Author_Institution
Coll. of Food Sci. & Technol., Shanghai Ocean Univ., Shanghai, China
Volume
9
fYear
2010
fDate
16-18 Oct. 2010
Firstpage
4217
Lastpage
4220
Abstract
The thermal probe method can be conveniently and rapidly used to measure the thermal conductivities of liquids. But the bridge voltages chosen differently would have large effect on output electric voltage signal ΔV which would directly influence the results of experiments. To obtain the best value of voltage, the effects of different bridge voltages on thermal conductivities of samples were studied by experiments and theoretical analysis of heat transfer and numerical simulation. The results showed that the deviation of thermal conductivities were below 2 % when the bridge voltage was at 2.0V~8.0V. The number of GrPr was 1×103~0.98×104 which was less than 1×104. Each measurement was completed within one minute and the temperature rise was less than under 1 °C. It can effectively reduce the effect of convection and radiation to obtain accurately the thermal conductivity of liquid.
Keywords
bridge instruments; thermal conductivity; thermal variables measurement; bridge voltage; convection; heat transfer; liquid thermal conductivity measurement; numerical simulation; radiation; thermal probe method; voltage 2.0 V to 8.0 V; Bridge circuits; Conductivity; Conductivity measurement; Probes; Temperature measurement; Thermal conductivity; Voltage measurement; bridge voltage; liquid; output electric voltage signal; thermal conductivity; thermal probe;
fLanguage
English
Publisher
ieee
Conference_Titel
Image and Signal Processing (CISP), 2010 3rd International Congress on
Conference_Location
Yantai
Print_ISBN
978-1-4244-6513-2
Type
conf
DOI
10.1109/CISP.2010.5646821
Filename
5646821
Link To Document