Title : 
Development of the digitized pressure difference meter
         
        
            Author : 
Zhu, Shi-Ming ; Lu, Jie ; Liang, Jun-Ting ; Qiang, Menglu
         
        
            Author_Institution : 
Inst. of Acoust., Tongji Univ., Shanghai, China
         
        
        
        
        
        
            Abstract : 
The principles, construction, error analysis and insite measurement results of the digitized pressure difference meter are presented. It measures gas pressure differences by measuring by ultrasonic means the depths of columns of liquid. Its accuracy measured by the China National Institute of Metrology in the range of 0-120 mm is 0.03%
         
        
            Keywords : 
computerised instrumentation; error analysis; pressure measurement; pressure sensors; ultrasonic measurement; 0 to 1173.99 Pa; 0 to 120 mm; digitized pressure difference meter; error analysis; gas pressure differences; liquid column depth; ultrasonic measurement; Acoustic transducers; Computer displays; Delay effects; Detectors; Equations; Microprocessors; Pressure measurement; Pulse circuits; Time measurement; Voltage;
         
        
        
        
            Conference_Titel : 
Industrial Technology, 1996. (ICIT '96), Proceedings of The IEEE International Conference on
         
        
            Conference_Location : 
Shanghai
         
        
            Print_ISBN : 
0-7803-3104-4
         
        
        
            DOI : 
10.1109/ICIT.1996.601611