DocumentCode :
3281783
Title :
Online dropout detection in subcutaneously implanted continuous glucose monitoring
Author :
Quan Shen ; Qin, S.J. ; Doniger, K.J.
Author_Institution :
Dept. of Chem. Eng., Univ. of Texas at Austin, Austin, TX, USA
fYear :
2010
fDate :
June 30 2010-July 2 2010
Firstpage :
4373
Lastpage :
4378
Abstract :
Continuous glucose monitoring (CGM) can inform diabetic people of the status of their glucose and warn of actual or impending hypo- or hyperglycemia. In the past decades, biosensors have been developed for CGM, among which the subcutaneously implanted glucose monitors are of much interest because they are less painful while quickly measure the critical glucose concentrations in the body. Experiments of these monitors show that there are occasionally spurious dropouts which do not reflect the true glucose level and will cause problems in glucose calibration. In this paper, a discrete wavelet transform (DWT) based online detector is proposed to detect dropouts for a single glucose sensor signal. Its effectiveness is demonstrated by both experimental data from the non-diabetic pig and clinical data from people.
Keywords :
biomedical measurement; discrete wavelet transforms; patient monitoring; DWT; biosensors; continuous glucose monitoring; diabetes; discrete wavelet transform; implants; online dropout detection; Biosensors; Blood; Calibration; Chemical engineering; Detectors; Diabetes; Discrete wavelet transforms; Monitoring; Sugar; User-generated content;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
American Control Conference (ACC), 2010
Conference_Location :
Baltimore, MD
ISSN :
0743-1619
Print_ISBN :
978-1-4244-7426-4
Type :
conf
DOI :
10.1109/ACC.2010.5530788
Filename :
5530788
Link To Document :
بازگشت