DocumentCode
3290800
Title
Piezoelectric shock wave generator with electronic focusing
Author
Cathignol, D. ; Chapelon, J.Y. ; Birer, A. ; Nachef, S.
Author_Institution
INSERM, Lyon, France
fYear
1991
fDate
8-11 Dec 1991
Firstpage
1335
Abstract
Electrohydraulic, electromagnetic and piezoelectric shock wave generators used in the treatment of renal calculi and gallstones have their focal zone determined by their geometrical parameters. A piezoelectric shock wave generator with electronic focalization is proposed. The system is composed of a bidimensional array of 25 piezoelectric independent transducers arranged in a spherical shell 10 cm in diameter. The electronic circuitry of the system includes 25 pulsers. The interdelay of each channel can be adjusted between 10 ns and 100 μs by steps of 10 ns. The pressure at the surface of the shell was 22×105 Pa. The focal point can be electronically moved in an ellipsoidal zone 3 cm in diameter and 4 cm long. These values are large enough to follow the displacement of the renal calculus during respiration
Keywords
biomedical electronics; biomedical ultrasonics; piezoelectric transducers; radiation therapy; shock wave effects; 10 cm; 10 ns to 100 mus; 2.2×106 Pa; 3 cm; 4 cm; bidimensional array; electronic focusing; gallstones treatment; piezoelectric shock wave generators; renal calculi treatment; transducers; Calculus; Composite materials; Delay; Electric shock; Electrohydraulics; Electromagnetic scattering; Kidney stones; Phased arrays; Shock waves; Ultrasonic transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 1991. Proceedings., IEEE 1991
Conference_Location
Orlando, FL
Type
conf
DOI
10.1109/ULTSYM.1991.234046
Filename
234046
Link To Document