DocumentCode
3291054
Title
HIP-KNEE control for gait assistance with Powered Knee Orthosis
Author
Wai-Yin Lai ; Hao Ma ; Wei-Hsin Liao ; Fong, Daniel Tik-Pui ; Kai-Ming Chan
Author_Institution
Dept. of Mech. & Autom. Eng., Chinese Univ. of Hong Kong, Hong Kong, China
fYear
2013
fDate
12-14 Dec. 2013
Firstpage
762
Lastpage
767
Abstract
A Powered Knee Orthosis (PKO) was developed for the elderly and patients with disordered gait to regain normal walking. In order to enhance the PKO performance and reduce system complexity especially for people with muscle weakness in their knee joints, an algorithm named HIP-KNEE control is proposed. This algorithm is based on the analysis of kinematic gait model, and the desired knee joint angle (KNEE) is estimated from the measurements of hip joint angle (HIP). The relationship between HIP and KNEE is modeled as a polynomial, which can be easily implemented to an embedded controller for real-time control. This control method is suitable to subjects with good function in hip joint, and it can provide help in walking without special training. An Inertia Measurement Units (IMU) is used for obtaining HIP input, and integrated with a footswitch for checking the heel condition; the gait assistance performance can be further improved.
Keywords
gait analysis; orthotics; polynomials; PKO performance; control method; disordered gait patients; elderly; embedded controller; footswitch; gait assistance; heel condition; hip joint angle; hip-knee control algorithm; inertia measurement unit; kinematic gait model; knee joint angle; normal walking; polynomial model; powered knee orthosis; realtime control; system complexity; Emulation; Hip; Joints; Knee; Legged locomotion; Testing; Torque;
fLanguage
English
Publisher
ieee
Conference_Titel
Robotics and Biomimetics (ROBIO), 2013 IEEE International Conference on
Conference_Location
Shenzhen
Type
conf
DOI
10.1109/ROBIO.2013.6739554
Filename
6739554