Title
Normalization factors for the ClearPET/spl trade/ Neuro
Author
Weber, Simone ; Gundlich, Brigitte ; Khodaverdi, Maryam
Author_Institution
Central Inst. for Electron., Forschungszentrum Julich GmbH
Volume
5
fYear
2005
fDate
23-29 Oct. 2005
Firstpage
2632
Lastpage
2635
Abstract
Inhomogeneities in the lines of response (LOR) sensitivity of a PET scanner are a source of artifacts in the reconstructed image. Appropriate normalization before reconstruction is therefore required to obtain artifact-free images. Differences in the LOR sensitivities are mainly due to different detector responses and geometric effects. The relative importance of the different normalization factors is affected by the individual scanner geometry and potential scanner rotation. We discuss relevant normalization factors for the ClearPETtrade Neuro, a recently developed small-animal PET scanner, and their importance with respect to the unconventional geometry of the scanner. It is shown that homogeneous images can be obtained, provided that all normalization procedures are performed properly
Keywords
image reconstruction; medical image processing; positron emission tomography; ClearPET Neuro; artifact-free images; image reconstruction; inhomogeneities; lines of response sensitivity; normalization factors; potential scanner rotation; scanner geometry; Crystals; Data acquisition; Detectors; Geometry; Head; Image reconstruction; Open source software; Photomultipliers; Positron emission tomography; Spatial resolution;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2005 IEEE
Conference_Location
Fajardo
ISSN
1095-7863
Print_ISBN
0-7803-9221-3
Type
conf
DOI
10.1109/NSSMIC.2005.1596878
Filename
1596878
Link To Document