Title :
A pulse sequence for improving MRI images
Author :
Malik, Aamir Saeed ; Song, Taek Lyul
Author_Institution :
Dept. of Electron. Inf. Syst. Eng., Hanyang Univ., Ansan
Abstract :
Improving SNR as well as resolution simultaneously is a challenging problem in MRI. In normal EPI sequences, a long data acquisition time not only excessively attenuates the signal obtained from the sequence but also alters its resolution depending on the adjustment of the sequence. In this paper, we address this issue by designing a pulse sequence for GE-EPI (Gradient Echo - Echo Planar Imaging). Instead of starting acquisition from the maximum point in K-Space, the acquisition point is selected according to the requirement. For example, if good SNR is required than the acquisition point selected is close to zero so as to preserve the DC component.
Keywords :
biomedical MRI; biomedical NMR; medical signal processing; spin echo (NMR); GE-EPI; MRI SNR improvement; MRI resolution improvement; biomedical MRI pulse sequence; gradient echo-echo planar imaging; Attenuation; Bandwidth; Biological tissues; Data acquisition; High-resolution imaging; Image resolution; Magnetic resonance imaging; Magnetization; Signal resolution; Signal to noise ratio;