DocumentCode
3303875
Title
Shear activation of platelets in an axial flow ventricular assist device
Author
Snyder, T.A. ; Litwak, K.N. ; Watach, M.J. ; Wagner, W.R.
Author_Institution
Dept. of Bioeng., Pittsburgh Univ., PA, USA
Volume
2
fYear
1999
fDate
36434
Abstract
A number of axial flow ventricular assist devices (VAD) are currently undergoing large animal trials prior to clinical evaluation. The potential for shear activation of platelets by these turbine containing devices is of concern. In this study increases in platelet activation and consumption were demonstrated and quantified in response to increased shear stress levels within an experimental VAD
Keywords
artificial organs; cardiology; cellular biophysics; haemorheology; animal trials; axial flow ventricular assist device; clinical evaluation; flow cytometry; increased shear stress levels; platelets; shear activation; turbine containing devices; Animals; Stress; Turbines;
fLanguage
English
Publisher
ieee
Conference_Titel
[Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
Conference_Location
Atlanta, GA
ISSN
1094-687X
Print_ISBN
0-7803-5674-8
Type
conf
DOI
10.1109/IEMBS.1999.803908
Filename
803908
Link To Document