• DocumentCode
    3303977
  • Title

    Aortic pressure with asynchronous ventricular assist

  • Author

    Gu, M. ; Gaumond, R. ; Snyder, A. ; Gardner, J.

  • Author_Institution
    Dept. of Bioeng., Pennsylvania State Univ., University Park, PA, USA
  • Volume
    2
  • fYear
    1999
  • fDate
    36434
  • Abstract
    The authors have examined the relation between Pao (aortic pressure) and parameters of an asynchronous VAD (ventricular assist device) in a lumped-parameter simulation. While left ventricular volume was greater than a threshold level Vth, the 50 ml VAD chamber filled passively at 100 ml/sec. The VAD emptied at a rate of 400 ml/sec when filled. The ventricular elastance (1/C) followed a half-rectified sine waveform. Over a wide range of heart rates determined by Vth, VAD pumping was entrained synchronous to ventricular pumping by this protocol. With the VAD included, Pao(t) waveform complexity increased, with generally more high-frequency signal energy. This has implications for baroreceptor reflex function in VAD patients
  • Keywords
    artificial organs; cardiology; haemodynamics; physiological models; aortic pressure; asynchronous ventricular assist; baroreceptor reflex function; cardiovascular model; half-rectified sine waveform; high-frequency signal energy; left ventricular volume; lumped-parameter simulation; passive chamber filling; ventricular elastance; Baroreflex; Heart rate; Protocols;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    [Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
  • Conference_Location
    Atlanta, GA
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5674-8
  • Type

    conf

  • DOI
    10.1109/IEMBS.1999.803913
  • Filename
    803913