    Experimental assessment of error sources in thermodilution measurements of cardiac output and ejection fraction

  • Author

    Gefen, Nurit ; Barnea, Ofer ; Abramovich, Amir ; Santamore, W.P.

  • Author_Institution
    Dept. of Biomed. Eng., Tel Aviv Univ., Israel
  • Volume
    2
  • fYear
    1999
  • fDate
    36434
  • Abstract
    The most common method used for clinical measurement of cardiac output and right ventricular ejection fraction is thermodilution. This method is based on several assumptions, which are not always valid. This may result in errors in order of 20-50%. In this work these sources of error were experimentally examined. Results indicate that the errors associated with ejection fraction are mostly due to residual thermal energy at the right atrium that is not considered in the calculations. Cardiac output measurement suffers mostly from thermal losses along the catheter
  • Keywords
    biomedical measurement; biothermics; cardiology; heat losses; measurement errors; temperature measurement; cardiac output; catheter thermal losses; clinical measurement; error sources; residual thermal energy; right atrium; right ventricular ejection fraction; temperature measurement; thermodilution measurements; Catheters; Equations; Solids; Temperature; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    [Engineering in Medicine and Biology, 1999. 21st Annual Conference and the 1999 Annual Fall Meetring of the Biomedical Engineering Society] BMES/EMBS Conference, 1999. Proceedings of the First Joint
  • Conference_Location
    Atlanta, GA
  • ISSN
    1094-687X
  • Print_ISBN
    0-7803-5674-8
  • Type

    conf

  • DOI
    10.1109/IEMBS.1999.803951
  • Filename
    803951