    Testing of Several CT Performance Phantoms for Medical Imaging

  • Author

    Li Dehong ; Li Xingdong ; Wu Huili

  • Author_Institution
    Div. of Metrol. in Ionizing Radiat. & Med., Nat. Inst. of Metrol., Beijing, China
  • fYear
    2011
  • fDate
    10-12 May 2011
  • Firstpage
    1
  • Lastpage
    2
  • Abstract
    In order to compare and evaluate the parameters of the modules for Computed Tomography (CT) number, and check the performance of high contrast resolution, low contrast resolution, image uniformity of CT performance phantoms, approximately one hundred CT performance phantoms were tested under the same clinical conditions. The experimental results indicate that there are variations testing results to the different CT performance phantoms in despite of using the same CT scanner. The results showed that in order to meet the requirements of CT performance test, CT performance phantom standard should be established and traceability.
  • Keywords
    computerised tomography; image resolution; medical image processing; phantoms; CT performance phantoms; CT scanner; computed tomography number; high contrast resolution; image uniformity; low contrast resolution; medical imaging; phantoms; Computed tomography; Image resolution; Materials; Phantoms; Testing; X-ray imaging;