DocumentCode :
3323162
Title :
Real-Time Detection of Loop-Mediated Isothermal Amplification Reaction on Microfluidic Chip
Author :
Xu, Ling-Yan ; Jin, Wei ; Zhu, Qiang-Yuan ; Yang, Wen-Xiu ; Wu, Qing-Qing ; Jin, Qin-Han ; Mu, Ying
Author_Institution :
State Key Lab. of Ind. Control Technol., Zhejiang Univ., Hangzhou, China
fYear :
2011
fDate :
10-12 May 2011
Firstpage :
1
Lastpage :
3
Abstract :
A device have been developed for monitoring the process of DNA loop-mediated isothermal amplification (LAMP) reaction through absorbance measurement. The LAMP reaction has a high efficiency and can synthesize large amounts of DNA in a short period of time. As its by-product pyrophosphate ions can be combined with magnesium ions, forming a white precipitate of magnesium pyrophosphate. The change of sample absorbance can be detected on line by an optic sensor. We can not only determine whether the LAMP reaction happened or not by the difference of absorbance between the start point and the end point but also monitor the dynamics process of LAMP reaction.
Keywords :
DNA; biochemistry; microfluidics; molecular biophysics; DNA loop-mediated isothermal amplification reaction; LAMP reaction; absorbance measurement; microfluidic chip; optic sensor; pyrophosphate ion; real-time detection; DNA; Isothermal processes; Microfluidics; Monitoring; Optical fiber sensors; Optical fibers; Real time systems;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioinformatics and Biomedical Engineering, (iCBBE) 2011 5th International Conference on
Conference_Location :
Wuhan
ISSN :
2151-7614
Print_ISBN :
978-1-4244-5088-6
Type :
conf
DOI :
10.1109/icbbe.2011.5780331
Filename :
5780331
Link To Document :
بازگشت