DocumentCode :
3326431
Title :
Investigation of a sub-millimeter resolution PET detector with depth of interaction encoding using digital SiPM single sided readout
Author :
Düppenbecker, Peter Michael ; Lodomez, Sarah ; Haagen, Rainald ; Marsden, Paul K. ; Schulz, Volkmar
Author_Institution :
Philips Res. Eur., Aachen, Germany
fYear :
2011
fDate :
23-29 Oct. 2011
Firstpage :
2252
Lastpage :
2253
Abstract :
One of the main challenges for building high resolution PET scanners with sub-millimeter resolution is to compensate for the parallax error induced by the height of the detector elements. During the last years several approaches to extract depth of interaction information for pixilated scintillation detectors were evaluated. We investigate the feasibility of a two layer PET detector with single sided readout using digital SiPM (dSiPM) arrays whereby the depth information is encoded in the width of the light distribution.
Keywords :
nuclear electronics; positron emission tomography; readout electronics; scintillation counters; silicon radiation detectors; PET detector; depth-of-interaction encoding; depth-of-interaction information; detector elements; digital SiPM arrays; high resolution PET scanners; light distribution width; parallax error; pixilated scintillation detectors; single sided readout; sub-millimeter resolution; Photonics;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
ISSN :
1082-3654
Print_ISBN :
978-1-4673-0118-3
Type :
conf
DOI :
10.1109/NSSMIC.2011.6152490
Filename :
6152490
Link To Document :