DocumentCode :
3327228
Title :
Analysis of the signal for high resolution digital X-ray detector under radiotherapy irradiation condition
Author :
Kim, Jong Eon ; Kim, Min Woo ; Kim, Yoon Suk ; Oh, Kyung Min ; Yun, Min Seok ; Kwon, Choul ; Nam, Kyung Tae ; Nam, Sang Hee
Author_Institution :
Radiol. Sci., Kaya Univ., Gimhae, South Korea
fYear :
2009
fDate :
Oct. 24 2009-Nov. 1 2009
Firstpage :
3291
Lastpage :
3295
Abstract :
In order to get a higher resolution of the image for a patient, the direct conversion mode which X-ray converts to electric charge are better than the indirect conversion mode which X-ray converts to light. But, the flat panel X-ray image detectors for direct conversion mode are not yet commercially available. In this study, we fabricated the samples for X-ray image film by using polycrystalline mercury iodide and carried out a feasibility study on the radiotherapy application. A sheet of copper (1 mm thickness) was attached onto the radiotherapy X-ray image film detector by insulating tape. The characteristics of the samples such as X-ray sensitivity, reproducibility, X-ray response on dose and dose rate for 6 MV beam was investigated. Dark currents were measured as a function of applied electric fields. The attenuated fraction of samples exposed with 6 and 10 MV beams was also measured as a function its thicknesses. By data analysis, the reproducibility of electric current measurements was 1.75% in the coefficient of variation. The measured values of electric charge revealed linearity with respect to the dose and dose rate, respectively. On the other hand, as the dose increases, the X-ray sensitivity decreased due to polarization effect. And the attenuated fraction for a 6 MV beam was 0.75% at the sample with the thickness of 380 ¿m. Therefore, the fabricated samples for X-ray image film provides the feasibility of radiotherapy imaging application. Moreover, it is found that the thickness of polycrystalline mercury iodide is very important factor for radiotherapy imaging, and a proper thickness having the attenuated fraction of about 1% should be greater than about 400 yW.
Keywords :
X-ray detection; dark conductivity; diagnostic radiography; image sensors; radiation therapy; X-ray response; X-ray sensitivity; attenuated fraction; dark current; dose rate; electric charge; electric current measurements; flat X-ray image film; flat panel X-ray image detector; high resolution digital X-ray detector; polycrystalline mercury iodide; radiotherapy irradiation condition; reproducibility; size 1 mm; size 380 mum; voltage 10 MV; voltage 6 MV; Current measurement; Electric variables measurement; Image converters; Optical imaging; Reproducibility of results; Signal analysis; Signal resolution; X-ray detection; X-ray detectors; X-ray imaging; Mercury iodide; Radiotherapy; direct conversion; flat panel detector;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
ISSN :
1095-7863
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2009.5401733
Filename :
5401733
Link To Document :
بازگشت