• DocumentCode
    3327692
  • Title

    CT-based attenuation correction on the FLEX Triumph™ preclinical PET/CT scanner

  • Author

    Prasad, Rameshwar ; Ay, Mohammad R. ; Ratib, Osman ; Zaidi, Habib

  • Author_Institution
    Div. of Nucl. Med., Geneva Univ. Hosp., Geneva, Switzerland
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    3357
  • Lastpage
    3362
  • Abstract
    Positron Emission Tomography (PET) has emerged as a very valuable molecular imaging modality for quantitative measurement of biochemical processes in the clinical and preclinical imaging domains. However, PET imaging suffers from various physical degrading factors including photon attenuation, which can be corrected using CT-based attenuation correction (CTAC) on combined PET/CT scanners. The attenuation map is calculated by converting CT numbers derived from low-energy polyenergetic x-ray spectra to linear attenuation coefficients at 511 keV. In this study, we implemented and evaluated tube voltage-dependent attenuation correction on the FLEX TriumphTM preclinical PET/CT scanner. The impact of both bilinear and quadratic polynomial calibration curve methods obtained at different CT tube voltages was assessed. Correlation coefficients and best fit equations were calculated for both techniques. Phantoms and rodents´ PET images were used to assess improvements in image quality and quantitative accuracy. It was observed that the slopes of the bilinear calibration curves for CT numbers greater than 0 HU increase with increasing tube voltage. Significantly higher correlation coefficients were obtained using the quadratic polynomial calibration curve method. Acquired phantoms and rodents´ PET data were successfully corrected for photon attenuation using the CTAC procedure thus enabling more accurate quantitative analysis of biochemical processes in vivo.
  • Keywords
    biochemistry; computerised tomography; medical image processing; phantoms; polynomial approximation; positron emission tomography; CT-based attenuation correction; FLEX Triumph preclinical PET-CT scanner; bilinear polynomial calibration curve method; biochemical process; correlation coefficients; electron volt energy 511 keV; low-energy polyenergetic X-ray spectra; molecular imaging modality; phantoms; photon attenuation; positron emission tomography; quadratic polynomial calibration curve method; Attenuation; Calibration; Computed tomography; Imaging phantoms; Molecular imaging; Optical imaging; Polynomials; Positron emission tomography; Rodents; Voltage;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5401757
  • Filename
    5401757