Title :
Measurements for the SiliPET project: A small animal PET scanner based on stacks of silicon detectors
Author :
Auricchio, N. ; Di Domenico, G. ; Milano, L. ; Malaguti, R. ; Ambrosi, G. ; Ionica, M. ; Fiandrini, E. ; Zavattini, G.
Author_Institution :
Dept. of Phys., Univ. of Ferrara, Ferrara, Italy
fDate :
Oct. 24 2009-Nov. 1 2009
Abstract :
In this paper we propose a new scanner for small-animal positron emission tomography (PET) based on stacks of double sided silicon detectors. Each stack is composed of 40 planar detectors with dimension 60 à 60 à 1 mm3 and 128 orthogonal strips on both sides to read the two coordinates of interaction, the third being the detector number in the stack. Multiple interactions in a stack are discarded by an exclusive OR applied between each plane detector of a stack. In this way we achieve a precise determination of the interaction point of the two 511 keV photons. The reduced dimensions of the scanner also improve the solid angle coverage resulting in a high sensitivity. Preliminary results were obtained with MEGA prototype tracker (11 double sided Si detector layers, each with a thickness of 0.5 mm and a strip pitch of 470 ¿m), divided into two stacks 2 cm apart made of respectively 5 and 6 prototype layers, placing a small spherical 22Na source in different positions. We report on the results, spatial resolution, imaging, spectral and timing performances obtained with double sided silicon detectors, manufactured by ITC-FBK, having an active area of 3 à 3 cm2 and a strip pitch of 500 ¿m. Two different strip widths of 300 ¿m and 200 ¿m, and two thicknesses of 1 mm and 1.5 mm, equipped with 64 orthogonal p and n strips on opposite sides were read out with the VATAGP2.5 ASIC, a 128-channel ¿general purpose¿ charge sensitive amplifier.
Keywords :
application specific integrated circuits; position sensitive particle detectors; positron emission tomography; silicon radiation detectors; 128-channel charge sensitive amplifier; MEGA prototype tracker; VATAGP2.5 ASIC; double sided silicon detectors; electron volt energy 511 keV; imaging performance; orthogonal strips; planar detectors; size 0.5 mm; size 1 mm; size 1.5 mm; size 200 mum; size 300 mum; small animal PET scanner; small-animal positron emission tomography; solid angle coverage; spatial resolution; spectral performance; timing performance; Animals; Detectors; Manufacturing; Positron emission tomography; Prototypes; Silicon; Solids; Spatial resolution; Strips; Timing;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2009.5401809