Evaluation of a fully 3D, big bore TOF PET scanner with reduced scatter shields
Author :
Scheuermann, Joshua S. ; Surti, Suleman ; Kolthammer, Jeffrey A. ; Karp, Joel S.
Author_Institution :
Dept. of Radiol., Univ. of Pennsylvania, Philadelphia, PA, USA
fYear :
2009
fDate :
Oct. 24 2009-Nov. 1 2009
Firstpage :
3589
Lastpage :
3592
Abstract :
Traditionally, PET scanners have annular lead shielding at the axial ends that extends beyond the crystals to reduce the number of photons from outside of the axial field of view (FOV) hitting the detector. In recent years there has been a trend toward reducing the end shielding in order to increase the patient port diameter of the PET scanner. The reduction of the lead shielding could have performance effects that could affect overall image quality. The University of Pennsylvania has both a Philips Gemini TF Big Bore and a standard Gemini TF, with the major difference between the two systems being the reduction of the end shielding on BigBore. We evaluated the count-rate performance between the two systems to determine differences in performance characteristics. We also performed phantom measurements to determine the impact of performance differences on the scatter correction and reconstructed image quality. While there are differences in the performance of BigBore as compared to TF, the overall image quality of studies obtained on BigBore are comparable to those obtained on TF.
Keywords :
image reconstruction; medical image processing; phantoms; positron emission tomography; shielding; BigBore; Philips Gemini TF Big Bore; University of Pennsylvania; annular lead shielding; axial field of view; count-rate performance; fully 3D big bore TOF PET scanner; phantom measurements; reconstructed image quality; reduced scatter shield; scatter correction; standard Gemini TF; Boring; Detectors; Electromagnetic scattering; Image quality; Image reconstruction; Imaging phantoms; Particle scattering; Performance evaluation; Photonic crystals; Positron emission tomography;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
ISSN :
1095-7863
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2009.5401826
Filename :
5401826
Link To Document :
بازگشت