Title :
Monte Carlo PET camera modelling for proton range evaluation in proton therapy
Author :
Van Ngoc Ty, Claire ; De Marzi, L. ; Jan, Sebastien ; Lestand, Loic ; Ferrand, Régis ; Comtat, Claude ; Trébossen, Régine
Author_Institution :
I2BM/SHFJ/LPAIM, CEA, Orsay, France
fYear :
2011
fDate :
23-29 Oct. 2011
Firstpage :
3501
Lastpage :
3504
Abstract :
Proton therapy is a conformal method for dose delivery in sensitive regions. Offline PET monitoring has been proposed as proton irradiation produces positron emitters such as 11C and 15O [1]. Generally, the proposed methods consist in comparing a Monte Carlo simulation of the positron emitters produced during irradiation with PET measurements. Simulations can be performed using FLUKA [2] or GEANT4 [3] toolkit. Simulations are then convolved by a Gaussian function to account for the PET acquisition process. Such method assumes that the point spread function is constant within the tomograph field of view and that the PET camera is detecting all decay events. However the PET sensitivity is generally close to few percent. The aim is to investigate the contribution of the PET camera modelling either using a Monte-Carlo modelling of the PET camera with GATE [4] toolkit or a Gaussian post filtering of the distribution of the positron emitters for proton monitoring using PET. Experimental data were obtained from two single Bragg peaks in plexiglass. As expected the image statistics differs between the two camera modelling. However the range difference estimation observed was the same in homogeneous phantoms.
Keywords :
Gaussian processes; Monte Carlo methods; cameras; dosimetry; positron emission tomography; radiation therapy; FLUKA simulation; GEANT4 simulation; Gaussian post filtering; Monte Carlo PET camera modelling; dose delivery; image statistics; plexiglass; positron emitters; proton irradiation; proton monitoring; proton range evaluation; proton therapy; single Bragg peaks; toolkit simulation; Biological system modeling; Biomedical monitoring; Cameras; Logic gates; Monitoring; Positron emission tomography; Three dimensional displays;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
ISSN :
1082-3654
Print_ISBN :
978-1-4673-0118-3
Type :
conf
DOI :
10.1109/NSSMIC.2011.6152643
Filename :
6152643
Link To Document :
بازگشت