• DocumentCode
    3329260
  • Title

    A comparison of the performance of high QE photomultiplier tubes to conventional photomultiplier tubes

  • Author

    Liang, H. ; Oldendick, J. ; Li, Y. ; Ordonez, C.E. ; Chang, W.

  • Author_Institution
    Med. Center, Dept. of Diagnostic Radiol., Rush Univ., Chicago, IL, USA
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    3615
  • Lastpage
    3617
  • Abstract
    We have carried out experiments to assess the performance advantage of high quantum efficiency (HQE) photomultiplier tubes (PMTs) over conventional PMTs for possible use in a cardiac SPECT system. In particular, we have conducted tests of pulse height resolution and position sensitivity of HQE PMTs and compared their performance to conventional PMTs. Furthermore, we have conducted a cathode sensitivity scan and pulse height resolution scan of four HQE PMTs and four conventional PMTs. The HQE PMTs outperformed the conventional PMTs in both pulse height resolution and position sensitivity. Finally, we tested the four HQE PMTs on a 3 mm × 3 mm pixellated NaI(Tl) detector module and were able to resolve the pixels with 8.3% average energy resolution.
  • Keywords
    photomultipliers; pulse height analysers; scintillation counters; single photon emission computed tomography; NaI(Tl) detector module; cardiac SPECT system; cathode sensitivity scan; high quantum efficiency photomultiplier tubes; position sensitivity; pulse height resolution scan; size 3 mm; Photomultipliers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5401835
  • Filename
    5401835