Title :
Spatial resolution of the HRRT PET scanner using 3D-OSEM PSF reconstruction
Author :
Olesen, Oline Vinter ; Sibomana, Merence ; Keller, Sune Hogild ; Andersen, Flemming ; Jensen, Jorgen ; Holm, Soren ; Svarer, Claus ; Hjgaard, L.
fDate :
Oct. 24 2009-Nov. 1 2009
Abstract :
In this paper, the resolution of the Siemens high resolution research tomograph (HRRT) was centrally (r < 60 mm) homogenous with a FWHM of 1.4 mm for 18F-FDG in air. This was where the main part of the brain is located if the patient has been positioned correctly. The 1.4 mm resolution was obtained using the newly develop 3D-OSEM PSF reconstruction algorithm, which was a significant improvement over 3D-OSEM reconstruction without PSF. The algorithm uses a simple PSF model that was the same for all the pixels in the FOV and does not regulate for the circular/octagonal scanner geometry. This supports that the FWHM of the radial axis is increasing with the distance from the center for r > 60mm.
Keywords :
image reconstruction; image resolution; medical image processing; positron emission tomography; 3D-OSEM PSF reconstruction; 18F-FDG; HRRT PET scanner; Siemens high-resolution research tomograph; brain; circular scanner geometry; octagonal scanner geometry; size 1.4 mm; spatial resolution; Positron emission tomography; Spatial resolution;