Title :
Component-based normalization for the HRRT for sinogram-mode reconstruction
Author :
Sibomana, Merence ; Keller, Sune Hogild ; Holm, Soren ; Bloomfield, Peter M. ; Blinder, Stephan ; Hansen, Soren Baarsgaard ; Michel, Christian
fDate :
Oct. 24 2009-Nov. 1 2009
Abstract :
THE Siemens HRRT scanner is a complex PET tomograph, using 8 LSO/LYSO panel detectors and Depth Of Interaction(DOI) with a total of 119908 detector crystals. The Direct Normalization (DN) is the standard method used for sinogram-mode reconstruction. This paper investigates the use of an improved component-based normalization method for HRRT sinogram reconstruction. This method is originally used for MOLAR LOR listmode reconstruction.Three effects have been identified in the implementation of the original method that caused artifacts and quantification bias in the reconstructed uniformity image. The new CBN method has been implemented that corrects for these effects and produces a uniformity image with the same quality as the DN method. The main effect was the sensitivity drop for LORs with small angles of incidence, which affects all scanners by design.
Keywords :
image reconstruction; medical image processing; positron emission tomography; CBN method; LOR; PET; Siemens HRRT scanner; artifacts; component-based normalization; depth of interaction; direct normalization; line of response; quantification bias; sinogram-mode reconstruction; Crystals; Detectors; Helium; Image quality; Image reconstruction; Imaging phantoms; Noise level; Nuclear and plasma sciences; Positron emission tomography; Software testing;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2009.5401983