DocumentCode :
3332427
Title :
Flow Injection Spectrophotometric Determination of Cetylpyridinium Bromide in Environmental Water Samples
Author :
Li Hua
Author_Institution :
Key Subject Lab. for Radioactive Waste & Environ. Security, South West Univ. of Sci. & Technol., Mianyang, China
fYear :
2011
fDate :
10-12 May 2011
Firstpage :
1
Lastpage :
4
Abstract :
A flow injection spectrophotometric method for rapid determination of Cetylpyridinium Bromide (CPB) is proposed in this paper. The method is based on the reaction of Cetylpyridinium Bromide with xylenol orange (XO) in the medium of (CH2)6N4-HCl (pH=5.4) buffer solution. The absorbance intensity was registered in this reaction solution at 520 nm. The calibration graph is linear in the range of 0.5-2.4mg I-1. The detection limit is 0.11 mg I-1 at a sampling frequency of 20h-1. The relative standard deviation was 0.43% CPB (n=6). This method is very simple, rapid and suitable for automatic and continuous analysis. The presented system has been applied successfully to determination of CPB of environmental water samples.
Keywords :
organic compounds; pH; spectrochemical analysis; spectrophotometry; surfactants; visible spectra; water pollution measurement; absorbance intensity; automatic analysis; calibration graph; cationic surfactant; cetylpyridinium bromide; continuous analysis; environmental water samples; flow injection spectrophotometric determination; reaction solution; relative standard deviation; wavelength 520 nm; xylenol orange; Absorption; Calibration; Chemicals; Coils; Optimization; Valves; Water pollution;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Bioinformatics and Biomedical Engineering, (iCBBE) 2011 5th International Conference on
Conference_Location :
Wuhan
ISSN :
2151-7614
Print_ISBN :
978-1-4244-5088-6
Type :
conf
DOI :
10.1109/icbbe.2011.5780797
Filename :
5780797
Link To Document :