Title :
Feasibility of a high-resolution pixellated detector for C-SPECT
Author :
Sabet, Hamid ; Liang, Haoning ; Li, Yusheng ; Rozler, Michael ; Oldendick, James ; Chang, Wei
Author_Institution :
Med. Center, Dept. of Diagnostic Radiol., Rush Univ., Chicago, IL, USA
fDate :
Oct. 24 2009-Nov. 1 2009
Abstract :
We have investigated the feasibility of utilizing high resolution pixellated NaI(Tl) detectors for C-SPECT, a high performance imaging system to be designed for cardiac applications. Conventional 1.5¿ single-anode PMTs were used as light detection devices. A detector module with 2.9 à 2.9 à 5 mm3 pixels (Saint-Gobain) was evaluated in terms of decoding map quality and energy resolution. Furthermore, two 3.2 à 3.2 à 10 mm3 in-house modules with Teflon and Lumirror were built to investigate higher sensitivity detectors and different reflector materials.
Keywords :
anodes; collimators; position sensitive particle detectors; single photon emission computed tomography; C-SPECT; Saint-Gobain pixellated detector; decoding map quality; energy resolution; high-resolution pixellated detector; higher sensitivity detectors; light detection devices; single-anode; Collimators; Decoding; Detectors; Energy resolution; Glass; Heart; High-resolution imaging; Image edge detection; Pixel; Windows;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2009.5402010