Title :
Improvement of HPGe detectors timing using pulse shape analysis
Author :
Crespi, F.C.L. ; Vandone, V. ; Brambilla, S. ; Camera, F. ; Million, B. ; Riboldi, S. ; Wieland, O.
Author_Institution :
INFN sez. di Milano, Univ. di Milano, Milan, Italy
fDate :
Oct. 24 2009-Nov. 1 2009
Abstract :
This work aims to investigate the improvement the HPGe detector time resolution using signal shape analysis . For this purpose a measurement using a coaxial HPGe detector, a LaBr3 scintillator and 60Co gamma source was taken. The RS Pulse Shape Analysis algorithm was then applied to the experimental signals for extracting time information and the results obtained were compared with those obtained using a standard CFD. The results show a 50% improvement in time resolution (4 ns FWHM) when the RS algorithm is used instead of a standard CFD (8 ns FWHM) algorithm.
Keywords :
germanium radiation detectors; nuclear electronics; pulse shaping; scintillation counters; 60Co gamma source; LaBr3 scintillator; coaxial HPGe detectors; high-purity germanium detectors; pulse shape analysis; signal shape analysis techniques; standard constant fraction discriminator; time information; Coaxial components; Computational fluid dynamics; Gamma ray detection; Gamma ray detectors; Gas detectors; Pulse shaping methods; Shape; Signal analysis; Signal resolution; Timing;
Conference_Titel :
Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
Conference_Location :
Orlando, FL
Print_ISBN :
978-1-4244-3961-4
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2009.5402106