• DocumentCode
    3334504
  • Title

    Improvement of HPGe detectors timing using pulse shape analysis

  • Author

    Crespi, F.C.L. ; Vandone, V. ; Brambilla, S. ; Camera, F. ; Million, B. ; Riboldi, S. ; Wieland, O.

  • Author_Institution
    INFN sez. di Milano, Univ. di Milano, Milan, Italy
  • fYear
    2009
  • fDate
    Oct. 24 2009-Nov. 1 2009
  • Firstpage
    2121
  • Lastpage
    2124
  • Abstract
    This work aims to investigate the improvement the HPGe detector time resolution using signal shape analysis . For this purpose a measurement using a coaxial HPGe detector, a LaBr3 scintillator and 60Co gamma source was taken. The RS Pulse Shape Analysis algorithm was then applied to the experimental signals for extracting time information and the results obtained were compared with those obtained using a standard CFD. The results show a 50% improvement in time resolution (4 ns FWHM) when the RS algorithm is used instead of a standard CFD (8 ns FWHM) algorithm.
  • Keywords
    germanium radiation detectors; nuclear electronics; pulse shaping; scintillation counters; 60Co gamma source; LaBr3 scintillator; coaxial HPGe detectors; high-purity germanium detectors; pulse shape analysis; signal shape analysis techniques; standard constant fraction discriminator; time information; Coaxial components; Computational fluid dynamics; Gamma ray detection; Gamma ray detectors; Gas detectors; Pulse shaping methods; Shape; Signal analysis; Signal resolution; Timing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5402106
  • Filename
    5402106