• DocumentCode
    333617
  • Title

    Reduction of NIRS near-surface sensitivity using two detectors

  • Author

    Himeno, Hiromi ; Fujii, Mamiko ; Suzuki, Akifumi ; Nakayama, Kiyoshi

  • Author_Institution
    Dept. of Electr. & Electron. Eng., Sophia Univ., Tokyo, Japan
  • Volume
    4
  • fYear
    1998
  • fDate
    29 Oct-1 Nov 1998
  • Firstpage
    1862
  • Abstract
    A 1-source and 1-detector system commonly used in NIRS (near infrared spectroscopy) has considerable sensitivity in the near-surface region, which makes the measurement sensitive to skin blood flow and the thickness of the intervening tissue. This report proposes a 1-source and 2-detectors system in which logarithmic output of the 2-detectors are subtracted, giving the suppressed sensitivity in the near-surface region
  • Keywords
    bio-optics; biological tissues; biomedical equipment; infrared spectroscopy; oximetry; photon transport theory; NIR spectroscopy; logarithmic output; model analysis; near-surface sensitivity reduction; oxygen consumption measurement; photon fluence density function; semi-infinite turbid media; skin blood flow sensitivity; suppressed sensitivity; tissue thickness sensitivity; two detectors use; Absorption; Blood flow; Detectors; Electromagnetic scattering; Equations; Fluid flow measurement; Infrared spectra; Particle scattering; Skin; Thickness measurement;