    We have developed a flexible x-ray micro-CT system, named FaCT, capable of changing its geometric configuration and acquisition protocol in order to best suit an object being imaged for a particular diagnostic task. High-performance computing technologies have been a major enabling factor for this adaptive CT system in terms of system control, fast reconstruction, and data analysis. In this work, we demonstrate an adaptive procedure in which a quick, sparse-projection pre-scan is performed, the data are reconstructed, and a region of interest is identified. Next, a diagnostic-quality scan is performed where, given the region of interest, the control computer calculates an illumination window for on-line control of an x-ray source masking aperture to transmit radiation only through the region of interest throughout the scan trajectory. Finally, the diagnostic scan data are reconstructed, with the region of interest being clearly resolved. We use a combination of a multi-core CPU and a pair of NVIDIA Tesla GPUs to perform these tasks.
  • Keywords
    computerised tomography; data analysis; medical image processing; FaCT; NVIDIA Tesla CPU; X-ray source masking aperture; acquisition protocol; adaptive CT system; controlled-dose imaging; data analysis; diagnostic-quality scan; flexible X-ray microCT system; geometric configuration; high-performance computing technology; high-resolution imaging; multicore CPU; on-line control; region-of-interest imaging; sparse-projection pre-scan; Adaptive control; Adaptive systems; Computed tomography; Computers; High-resolution imaging; Image reconstruction; Optical imaging; Programmable control; Protocols; X-ray imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record (NSS/MIC), 2009 IEEE
  • Conference_Location
    Orlando, FL
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-3961-4
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2009.5402313
  • Filename
    5402313