Incorporating count-rate dependence into model-based PET scatter estimation
Author
Stearns, Charles W. ; Manjeshwar, Ravindra M.
Author_Institution
PET Syst. Eng., GE Healthcare, Waukesha, WI, USA
fYear
2011
fDate
23-29 Oct. 2011
Firstpage
3745
Lastpage
3747
Abstract
Pile-up in a PET detector changes the sensitivity of the detector to unscattered and scattered annihilation photons differently. If the detection of a 511 keV photon is confounded by the arrival of a second photon during signal integration, it may be lost by failing the upper energy threshold, so its detection probability decreases as activity increases. If, however, the first photon is a scattered photon of energy less than the lower energy threshold, pile-up may lead to an energy signal that lies within the energy window; detection probability for this photon increases as activity increases. As a result, the measured scatter fraction will increase as activity increases in a PET scanner. This effect is readily observed in the NEMA NU 2-2007 Scatter Fraction and Count Losses test as an increase in measured scatter fraction, and a broadening of the scatter tails, as activity in the phantom increases. We incorporate this phenomenon into the model-based scatter estimate through a function, parameterized by a measurement of the detectors busy time, which modifies the detection probability matrix for the scattered and unscattered photons in the model. The function was determined by Monte Carlo simulation of the pileup performance of the block, and is therefore a function of the block configuration, the scintillator decay time and the signal integration time in the detector electronics, and the energy window used in the acquisition. The function is fit by a polynomial from 170 to 511 keV, the possible energy range for a single scattered annihilation photon. The resulting scatter model demonstrates a count rate dependence that better matches the scatter tails from the NEMA count rate experiment, and improves the quality of the scatter correction in certain high count rate patient studies.
Keywords
Monte Carlo methods; nuclear electronics; phantoms; positron emission tomography; probability; scintillation counters; Monte Carlo simulation; NEMA count rate experiment; PET scanner; count-rate dependence; detection probability matrix; detector electronics; electron volt energy 170 keV to 511 keV; model-based PET scatter estimation; phantom; scatter correction; scintillator decay time; signal integration time; single scattered annihilation photon;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location
Valencia
ISSN
1082-3654
Print_ISBN
978-1-4673-0118-3
Type
conf
DOI
10.1109/NSSMIC.2011.6153708
Filename
6153708
Link To Document