PET motion correction in LOR space using scanner-independent, adaptive projection data for image reconstruction with PRESTO
Author
Scheins, J.J. ; Ullisch, M. ; Tellmann, L. ; Weirich, Ch ; Kops, E. Rota ; Herzog, H. ; Shah, N.J.
Author_Institution
Inst. of Neurosci. & Med., Forschungszentrum Julich GmbH, Julich, Germany
fYear
2011
fDate
23-29 Oct. 2011
Firstpage
4099
Lastpage
4101
Abstract
Figure 2 shows reconstructed images of a two-chamber phantom (filled with two activity concentrations) which was measured in 7 different positions. The applied motion of the phantom induces a significant image degradation (left). Motion parameters were extracted from EPI volumes to correct the emission data before the reconstruction with PRESTO. The blurring due to motion completely vanishes (right).
Keywords
image reconstruction; medical image processing; motion compensation; phantoms; positron emission tomography; LOR space; PET motion correction; PRESTO; adaptive projection data; image degradation; image reconstruction; scanner independent; two chamber phantom; Additives;