Title :
Studies of electrode design for a sub-mm resolution 3-D position sensitive CZT PET detector
Author :
Gu, Yi ; Levin, Craig S.
Abstract :
Through a series of 6 studies, performance metrics of a CZT photon detector were experimentally measured as a function of electrode feature dimensions, viz. cathode strip width and pitch, and steering electrode width and bias potential. Along with 1 mm pitch anode strips, results indicate that a 5 mm cathode width and pitch, 400 μm steering electrode width, and -100 V steering electrode bias would best enhance detector performance.
Keywords :
anodes; biomedical electrodes; cathodes; photon counting; positron emission tomography; semiconductor counters; 3D position sensitive CZT PET detector; CZT photon detector; bias potential; cathode strip pitch; cathode strip width; electrode design; electrode feature dimension; steering electrode width; Application specific integrated circuits; Atmospheric measurements; Detectors; Electrodes; Particle measurements; Photonics; Spatial resolution;