Title
Readout technologies for the BNL-UPenn MRI-compatible PET scanner for rodents
Author
Purschke, Martin L. ; Budassi, M. ; Cao, T. ; Fried, J. ; Gualtieri, E. ; Junnarkar, S.S. ; Karp, J.S. ; Maramraju, S.-H. ; O´Connor, P. ; Pratte, J. -P ; Ravindranath, B. ; Schlyer, D. ; Stoll, S.P. ; Vaska, P. ; Woody, C.
Author_Institution
Brookhaven Nat. Lab., Upton, NY, USA
fYear
2011
fDate
23-29 Oct. 2011
Firstpage
617
Lastpage
620
Abstract
We describe the prototype of a full-body PET scanner for rats that is compatible with a 9.4 T MRI system. The detector consists of 96 PET detector blocks in a cylindrical arrangement. In this paper we concentrate on a new readout technology, which takes advantage of the fact that optical fibers are insensitive to electromagnetic interference. The data are formatted in a FPGA on the motherboard and sent to a data acquisition computer through standard Gigabit Ethernet connections. We will describe the technology chosen for the system, and introduce the data acquisition adapted for the readout of the data.
Keywords
biomedical MRI; biomedical electronics; field programmable gate arrays; positron emission tomography; readout electronics; BNL-UPenn MRI; FPGA; Gigabit Ethernet; PET scanner; data acquisition; electromagnetic interference; magnetic flux density 9.4 T; optical fibers; readout technologies; rodents; Crystals; Field programmable gate arrays; Heating; Image resolution; Magnetic resonance imaging;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location
Valencia
ISSN
1082-3654
Print_ISBN
978-1-4673-0118-3
Type
conf
DOI
10.1109/NSSMIC.2011.6154123
Filename
6154123
Link To Document