Title :
Comparison of microwave black-body target radiometric measurements
Author :
Walker, D.K. ; Gu, D. ; MacReynolds, K. ; Direen, R. ; Randa, J. ; Cox, A.E. ; Houtz, D. ; Billinger, R.L.
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
Abstract :
Accurate characterization of the brightness temperature (TB) of black-body targets used for calibrating microwave remote-sensing radiometers includes many inputs: antenna pattern and loss, target temperature, target emissivity, mechanical alignment, and radiometric TB measurements, all of which must be calibrated against physical standards. Here, we describe measurements made using several black-body targets and two different antennas within the WR-42 (18 to 26.5 GHz) waveguide band. Uncertainty estimates are also shown for the retrieved target TB measurements.
Keywords :
antenna radiation patterns; calibration; microwave measurement; radiometers; remote sensing; temperature measurement; WR-42 waveguide band; antenna loss; antenna pattern; frequency 18 GHz to 26.5 GHz; mechanical alignment; microwave black-body target radiometric measurement; microwave remote-sensing radiometer calibration; radiometric brightness temperature measurement; target emissivity; target temperature; Antenna measurements; Microwave antennas; Microwave measurements; Microwave radiometry; Radiometers; Temperature measurement; Radiometer calibration; calibration target; measurement uncertainty;
Conference_Titel :
Geoscience and Remote Sensing Symposium (IGARSS), 2010 IEEE International
Conference_Location :
Honolulu, HI
Print_ISBN :
978-1-4244-9565-8
Electronic_ISBN :
2153-6996
DOI :
10.1109/IGARSS.2010.5652481