DocumentCode :
3354469
Title :
High sensitive standard measurement to determine strength of an I-125 brachytherapy source
Author :
Unno, Yasuhiro ; Kurosawa, Tadahiro ; Yunoki, Akira ; Yamada, Takahiro ; Sato, Yasushi
Author_Institution :
Nat. Metrol. Inst. of Japan, AIST, Tsukuba, Japan
fYear :
2011
fDate :
23-29 Oct. 2011
Firstpage :
377
Lastpage :
381
Abstract :
We are encouraged to contribute to strengthen quality assurance which keeps brachytherapy safe and available. For the purpose, we aspire to determine the source strength and establish national metrological standard in Japan. It is necessary to assemble a high sensitive free air chamber for measurement of ionization current derived from the source which is very weak especially in Japan. Total uncertainty for the source strength is promising to meet required value from users of measurement instruments.
Keywords :
brachytherapy; iodine; ionisation chambers; measurement standards; radioactive sources; I; I-125 brachytherapy; free air chamber; ionization current; metrological standard; source strength; Apertures; Cable shielding; Energy measurement; IP networks; Instruments; Photonics; Pulleys;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
Conference_Location :
Valencia
ISSN :
1082-3654
Print_ISBN :
978-1-4673-0118-3
Type :
conf
DOI :
10.1109/NSSMIC.2011.6154522
Filename :
6154522
Link To Document :
بازگشت