• DocumentCode
    3354674
  • Title

    Optimization of pulse-height sharing for use below 150 keV in long bars of NaI(Tl)

  • Author

    Boyle, P.J. ; Hanna, D.S. ; MacLeod, A.M.L. ; Sinclair, L.E. ; Seywerd, H.C.J. ; Saull, P.R.B.

  • Author_Institution
    Phys. Dept., McGill Univ., Montreal, QC, Canada
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    440
  • Lastpage
    443
  • Abstract
    In this work we present an optimization of the energy and position resolutions for use in gamma-ray interactions in a long bar of NaI(TI). This work focuses on tuning the effective optical attenuation coefficient to approximately the same length of the bar. This is achieved by altering the surface of the scintillator in a systematic manner. We present data obtained in the laboratory from measurements of eighteen bars of NaI(TI) that have been manufactured with various treatments of the surface. This process results in an improvement in position resolution by a factor of four over a polished bar with only a corresponding 20% decrease in energy resolution.
  • Keywords
    gamma-ray detection; nuclear electronics; solid scintillation detectors; Nal(Tl); energy resolution; gamma-ray interaction; optical attenuation coefficient; position resolution; pulse-height sharing optimization; scintillator surface; Energy measurement; Image resolution; Optical imaging;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6154535
  • Filename
    6154535