• DocumentCode
    3357475
  • Title

    A novel ionization chamber for high doses at industrial gamma irradiation facilities

  • Author

    Ramirez-Jimenez, F.J. ; Pérez, Y. B Alcántara

  • Author_Institution
    Dept. de Sist. Electronicos, Inst. Nac. de Investig. Nucl., Ocoyoacac, Mexico
  • fYear
    2011
  • fDate
    23-29 Oct. 2011
  • Firstpage
    1811
  • Lastpage
    1813
  • Abstract
    This work presents the design of an ionization chamber with cylindrical geometry for measuring high doses of gamma radiation at industrial irradiation facilities. A novel geometry for the collector electrode to optimize the ion collection is suggested. This chamber is used to provide an alternative mean for dosimetry studies in the gamma irradiation plant, because with the proposed measuring system, we can know the dose rate at all the points and the dose during all the irradiation process. A prototype was tested and calibrated with 137Cs and 60Co radioactive sources in the Ionizing Radiation Metrology Laboratory of ININ and finally tested in the gamma irradiation plant at ININ.
  • Keywords
    dosimetry; electrodes; gamma-rays; geometry; ionisation chambers; radioactive sources; semiconductor counters; 137Cs; 60Co; ININ; calibration; collector electrode; cylindrical geometry; dosimetry study; gamma irradiation plant; industrial gamma irradiation facility; ion collection; ionizing radiation metrology laboratory; novel ionization chamber; radioactive sources; Ionization; Laboratories; Lead; Radiation effects;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
  • Conference_Location
    Valencia
  • ISSN
    1082-3654
  • Print_ISBN
    978-1-4673-0118-3
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2011.6154689
  • Filename
    6154689