Title : 
A novel ionization chamber for high doses at industrial gamma irradiation facilities
         
        
            Author : 
Ramirez-Jimenez, F.J. ; Pérez, Y. B Alcántara
         
        
            Author_Institution : 
Dept. de Sist. Electronicos, Inst. Nac. de Investig. Nucl., Ocoyoacac, Mexico
         
        
        
        
        
        
            Abstract : 
This work presents the design of an ionization chamber with cylindrical geometry for measuring high doses of gamma radiation at industrial irradiation facilities. A novel geometry for the collector electrode to optimize the ion collection is suggested. This chamber is used to provide an alternative mean for dosimetry studies in the gamma irradiation plant, because with the proposed measuring system, we can know the dose rate at all the points and the dose during all the irradiation process. A prototype was tested and calibrated with 137Cs and 60Co radioactive sources in the Ionizing Radiation Metrology Laboratory of ININ and finally tested in the gamma irradiation plant at ININ.
         
        
            Keywords : 
dosimetry; electrodes; gamma-rays; geometry; ionisation chambers; radioactive sources; semiconductor counters; 137Cs; 60Co; ININ; calibration; collector electrode; cylindrical geometry; dosimetry study; gamma irradiation plant; industrial gamma irradiation facility; ion collection; ionizing radiation metrology laboratory; novel ionization chamber; radioactive sources; Ionization; Laboratories; Lead; Radiation effects;
         
        
        
        
            Conference_Titel : 
Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2011 IEEE
         
        
            Conference_Location : 
Valencia
         
        
        
            Print_ISBN : 
978-1-4673-0118-3
         
        
        
            DOI : 
10.1109/NSSMIC.2011.6154689