• DocumentCode
    3359142
  • Title

    Study of 316L stainless steel coronary stent high vacuum heat treatment

  • Author

    Wang, Jia-Ling ; Chen, Zhuo ; Li, Yu-Tian

  • Author_Institution
    Changchun Inst. of Opt., Fine Mech. & Phys., Chinese Acad. of Sci., Changchun, China
  • fYear
    2009
  • fDate
    9-12 Aug. 2009
  • Firstpage
    2431
  • Lastpage
    2435
  • Abstract
    Establish the high vacuum heat treatment platform for high vacuum heat treatment of 316L stainless steel coronary stent; fix the art of high vaccum heat treatment through optimizing series parameters; by analysis the influence of heating-up temperature and holding time, draw the conclusion that we may inlet auxiliary gas and add wire netting to achieve bright heat treatment sample.
  • Keywords
    biomedical materials; heat treatment; stainless steel; stents; 316L stainless steel coronary stent; FeCr18Ni14Mo2.5; auxiliary gas; bright heat treatment sample; heating-up temperature; high vacuum heat treatment platform; holding time; series parameters; wire netting; Biological materials; Biomedical materials; Building materials; Furnaces; Heat pumps; Heat treatment; Steel; Temperature; Vacuum systems; Water heating; 316L stainless steel; coronary stent; high vacuum heat treatment;