DocumentCode :
336480
Title :
Accurate 3D pose estimation method for polyethylene wear assessment in total knee replacement
Author :
Fukuoka, Yutaka ; Hoshino, Akiho ; Ishida, Akimasa
Author_Institution :
Inst. for Med. & Dental Eng., Tokyo Med. & Dental Univ., Japan
Volume :
4
fYear :
1997
fDate :
30 Oct-2 Nov 1997
Firstpage :
1849
Abstract :
This study describes a new method for evaluating polyethylene wear in total knee prosthesis. Since the amount of wear is dependent on a number of variables such as weight and activity of the patient, the amount should be estimated based on in-vivo measurement. A computer vision technique called 3D/2D matching is employed to perform in-vivo assessment using single-plane radiograph. By the 3D/2D matching algorithm the three-dimensional position and orientation (pose) of knee implants is estimated to measure the femorotibial distance which is defined as the shortest perpendicular distance from the tibial tray to the femoral component. The accuracy of the proposed 3D/2D matching method is demonstrated on some in-vitro investigations. Pose estimation errors are examined and the maximum error in in-plane/out-of-plane translations and rotations 0.28/4.17 mm and 0.33/0.26 degrees, respectively. That in femorotibial distance measurements is 0.14 mm
Keywords :
biomechanics; biomedical materials; biomedical measurement; computer vision; diagnostic radiography; image matching; medical image processing; polymers; position measurement; prosthetics; wear; wear testing; 3D orientation; 3D pose estimation method; 3D/2D matching; computer vision technique; femoral component; femorotibial distance; in-plane/out-of-plane rotations; in-plane/out-of-plane translations; in-vitro investigations; in-vivo measurement; knee implants; polyethylene wear assessment; pose estimation errors; shortest perpendicular distance; single-plane radiograph; tibial tray; total knee prosthesis; total knee replacement; Computer vision; Diagnostic radiography; Distance measurement; Estimation error; Implants; In vitro; Knee; Polyethylene; Position measurement; Prosthetics;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1094-687X
Print_ISBN :
0-7803-4262-3
Type :
conf
DOI :
10.1109/IEMBS.1997.757093
Filename :
757093
Link To Document :
بازگشت