• DocumentCode
    3365331
  • Title

    A 2.6 MHz phased array ultrasonic probe using 0.67Pb(Mg13/Nb23/)O3-0.33PbTiO3 single crystal grown by the Bridgman method

  • Author

    Rhim, Sung Min ; Jung, Ho ; Kim, Seduk ; Lee, Sang-Goo

  • Author_Institution
    iBULe Humanscan Co Ltd, Ansan, South Korea
  • Volume
    2
  • fYear
    2002
  • fDate
    8-11 Oct. 2002
  • Firstpage
    1143
  • Abstract
    The performance of 64channel 2.6 MHz phased array ultrasonic probe using 0.67Pb(Mg13/ Nb23/)O3-0.33PbTiO3 single crystal was investigated. The fabrication process was modified as the mechanical and electrical properties difference between PMN-PT single crystal and conventional PZT. The sensitivity of the PMN-PT single crystal probe is about 6 dB higher than that of the conventional PZT probe, and the -6 dB bandwidth is about 30 % broader. The sensitivity variation of PMN-PT probe was within ±2.0 dB. The performance of PMN/PT probe with higher sensitivity was better than that of PZT probe in harmonic image. And the color Doppler of PMN-PT probe is superior than that of PZT probe. The PMN-PT single crystal provides improvement in the sensitivity and bandwidth.
  • Keywords
    biomedical ultrasonics; lead compounds; piezoceramics; ultrasonic arrays; ultrasonic imaging; 2.6 MHz; Bridgman method; PMN-PT single crystal probe; PbMgO3NbO3-PbTiO3; color Doppler; medical ultrasound imaging; phased array ultrasonic probe; Bandwidth; Ceramics; Crystals; Niobium; Optical imaging; Phased arrays; Probes; Temperature; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2002. Proceedings. 2002 IEEE
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-7582-3
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2002.1192496
  • Filename
    1192496