• DocumentCode
    3365969
  • Title

    Optimized flextensional transducers for an implantable hearing aid

  • Author

    Schmitt, Daniel ; Weber, Peter K. ; Pelinescu, Ion ; Park, Seung-Eek

  • Author_Institution
    Fraunhofer Inst. Biomedizinische Technik, St. Ingbert, Germany
  • Volume
    2
  • fYear
    2002
  • fDate
    8-11 Oct. 2002
  • Firstpage
    1273
  • Abstract
    Relaxor single crystal based flextensional transducers were investigated for an implantable hearing aid application. Oval shape cymbal structure, adopted due to anatomic requirement, was simulated using finite element analysis (FEA). The optimum lateral orientation was determined to be <100> along the longer direction of oval cap. The simulation results suggest that optimized single crystal transducers offer 100% improved performance over the analogous PZT transducers, adequate for implantable hearing aid device applications.
  • Keywords
    acoustic transducers; finite element analysis; hearing aids; piezoelectric transducers; relaxor ferroelectrics; anatomic requirement; finite element analysis; implantable hearing aid; optimized flextensional transducers; optimum lateral orientation; oval shape cymbal structure; relaxor single crystal; Acoustic transducers; Anisotropic magnetoresistance; Auditory system; Ceramics; Crystallography; Crystals; Niobium; Piezoelectric transducers; Shape; Vibrations;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2002. Proceedings. 2002 IEEE
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-7582-3
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2002.1192526
  • Filename
    1192526