DocumentCode :
336617
Title :
A new flexible receiver coil with a high signal-to-noise ratio for midfield MRI system
Author :
Khripov, A.A.
Author_Institution :
Dept. of Quantum Magnetic Phenomena, St. Petersburg State Univ., Russia
Volume :
2
fYear :
1997
fDate :
30 Oct-2 Nov 1997
Firstpage :
889
Abstract :
The distinctive property of the proposed flexible receiver coil consists in that it allows to change both the form and volume of the area enclosed in this coil. The sample occupies the whole volume of the coil. Signal to noise ratio is improved due to the reduction of the part of the electromagnetic power which is scattered by excited nuclear spin system in the surrounding environment. The technology of a proposed receiver coil manufacturing allows to vary the pitch and type of conductors. It allows to increase the loop area of the coil in order to reduce the resistance and improve quality factor of the receiver oscillating circuit which also improves signal to noise ratio
Keywords :
biomedical MRI; biomedical equipment; coils; excited nuclear spin system; flexible receiver coil; high signal-to-noise ratio; loop area; medical MRI; midfield MRI system; quality factor; receiver coil manufacture; receiver oscillating circuit; resistance reduction; varied conductor pitch; varied conductor type; Coils; Conductors; Electromagnetic scattering; Immune system; Magnetic fields; Magnetic resonance imaging; Magnetization; Nuclear magnetic resonance; Q factor; Signal to noise ratio;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 1997. Proceedings of the 19th Annual International Conference of the IEEE
Conference_Location :
Chicago, IL
ISSN :
1094-687X
Print_ISBN :
0-7803-4262-3
Type :
conf
DOI :
10.1109/IEMBS.1997.757810
Filename :
757810